| Sign In to gain access to subscriptions and/or personal tools. |
The Reliability of Doppler Ultrasound Techniques in the Assessment of Carotid DiseaseVascular Surgical Unit and Wessex Regional Medical Physics Department, Southampton, United Kingdom
Vascular Surgical Unit and Wessex Regional Medical Physics Department, Southampton, United Kingdom
Vascular Surgical Unit and Wessex Regional Medical Physics Department, Southampton, United Kingdom
Vascular Surgical Unit and Wessex Regional Medical Physics Department, Southampton, United Kingdom
Vascular Surgical Unit and Wessex Regional Medical Physics Department, Southampton, United Kingdom
The reliability of continuous wave (CW) Doppler imaging and Duplex scan ning in the assessment of carotid artery disease has been evaluated prospectively in comparison with biplanar angiography. Of 130 comparisons the sensitivity of Doppler imaging was 89% and specificity 99% in the detection of
CW Doppler imaging and Duplex scanning are accurate techniques in the diagnosis and categorisation of internal carotid artery disease. CW Doppler imaging is an effective screening technique for the presence of
Angiology, Vol. 36, No. 6,
333-340 (1985) This article has been cited by other articles:
|
|||||||||||||||
50% inter nal carotid stenosis, and 87% and 99% respectively in the diagnosis of internal carotid occlusion. Of 118 comparisons the sensitivity of Duplex scanning was 93% and specificity 98% in the detection of 